Sign In | Join Free | My tjskl.org.cn |
|
Categories | Walk-in Environmental Test Chamber |
---|---|
Brand Name: | Envsin |
Model Number: | ETCZ |
Certification: | CE and ISO |
Place of Origin: | China |
MOQ: | 1 |
Price: | To Be Negotiated |
Payment Terms: | L/C, T/T, Western Union |
Supply Ability: | 25 per month |
Delivery Time: | 60-90days after received the payment |
Packaging Details: | Wooden packing |
Product name: | Walk-in Temperature Humidity Chamber |
Key word: | Temperature Humidity |
Test Space: | 8m³/12m³/17m³/22m³/28m³ |
Internal size: | 2000*2000*2056mm (W*D*H) |
External size: | 2240mm*2240mm*3800mm (W*D*H) |
Cooling Method: | Water-cooled |
Temperature Fluctuation: | ±0.1~±0.5 |
Company Info. |
Envsin Instrument Equipment Co., Ltd. |
Verified Supplier |
View Contact Details |
Product List |
Walk-in Environmental Test Chamber with 10%~98% RH Humidity -40℃~180℃ Temperature
✔ Wide Temperature & Humidity Range: -40℃ to 180℃ with 10%~98% RH for versatile testing applications.
✔ Spacious Walk-in Design: Customizable interior for large or multiple test specimens.
✔ High-precision Control System: Ensures stable, uniform, and repeatable test conditions.
✔ Advanced Refrigeration Technology: Energy-efficient, reliable performance even at extreme
conditions.
✔ User-friendly Touchscreen Interface: Intuitive operation with programmable test settings.
✔ Safety & Protection Systems: Over-temperature, humidity, and electrical safety alarms for uninterrupted operation.
✔ Customizable to Your Needs: Tailored sizes, access ports, and configurations available.
Trusted by Global Industries | Compliant with International Standards | Durable & Low Maintenance
Temp. range | -40℃/-60℃~+80℃/100℃/150℃/+180℃ | Note: 1. Test with no load, meeting GB/T5170-2008, if no other special notice. 2. Ambient Temp 25℃, good ventilation condition: function well if 35℃≥RT>25℃;Cool down rate and thermal load capacity will be affected by ambient low temperature. |
Temp Fluctuation | ±0.1℃~±0.5℃ | |
Temperature Uniformity | ±0.5℃~±3.0℃ | |
Temp Deviation | <±2℃ | |
Heat up Time | <45min(R.T→+180℃) | |
Cool down Time | <90min(R.T→-70℃) | |
Humidity Test | Temp. Range: +10℃~+95℃ Humid. Range: 10%RH~98%RH Humid. Fluctuation: ±1.0%RH~±3.0%RH Humid. Uniformity ±1.0%RH~±5.0%RH Humid. Deviation: ±1.0%RH~±5.0%RH Temperature gradient: < 2.0 ℃ Humidity gradient: < 5.0%RH | |
Test Standards | 1.GB/T 2423.1 (Low Temp. Test) | 5.GJB 150.3 (High Temp. Test) |
2.GB/T 2423.2 (High Temp. Test) | 6.GJB 150.4 (Low Temp. Test) | |
3.GB/T2423.34 (Climate Cycling Test) | 7.GJB 150.9 (Climate Test) | |
4.IEC60068-2 (Climate Test) | 8.MIL-STD-202G-103B (Humid Test) | |
Load thermal capacity graph | In temperature test, the thermal capacity for loading varies from different temperature value. |
1.Chamber door | 2. Observation window |
3. Control Pad | 4. Working Lights |
5. Cable port | 6. Extensional control pad |
7. Wheels | 8. Power distribution board |
Structure: |
|
Construction: |
|
Load-bearin: |
|
Chamber door: |
|
Observation window: |
|
Cable ports to test spac: |
|
Lights: |
|
Shelves: |
|
Drain holes: |
|
Overflow hole: |
|
Wheels: |
|